You are here: System Overview > Dimension Icon Multi-Sample Chuck > Operation with the Multi-Sample Chuck > Small Samples

Small Samples

The magnet array of the Dimension Icon Multi-Sample Chuck is designed for working with small samples.

  1. Ensure that the Chuck Vacuum switch on the microscope is turned Off.
  2. Using “sticky dots” (supplied), glue your samples to the sample pucks.
  3. Ensure that the Multi-Sample Chuck is positioned with the 3 x 4 magnet array closest to the Dimension Icon head:

Figure 1: Chuck orientation relative to the scanner

  1. Place the loaded sample pucks on the magnet array.
  2. Plug the vacuum holes with the screws provided:

Figure 2: Plug the vacuum holes and load the sample pucks

  1. The sample puck centers are 40 mm (40,000 micrometers) apart in X and Y. Use this information for Programmed Moves or in Recipes.

 

www.bruker.com Bruker Corporation
www.brukerafmprobes.com 112 Robin Hill Rd.
nanoscaleworld.bruker-axs.com/nanoscaleworld/ Santa Barbara, CA 93117
   
  Customer Support: (800) 873-9750
  Copyright 2010, 2011. All Rights Reserved.